Pacific Scientific Instruments - Particle Counters
Pacific Scienticic Instruments - Met One

Particle Counting and the FDA

By Joe Gecsey, Product Manager

This article explores the FDA's approach to monitoring particle contamination and the need to protect particle counters in humid or washdown environments. The article was compiled frorn conversations with the FDA and operators of sterile fill lines.

THE IMPACT OF VARIOUS government regulations on actual manufacturing processes are not always obvious. The current government requirements for monitoring sterile fin lines in cleanrooms requires a specific method of particle counting that is somewhat different than Federal Standard 209D. 'Me particle counting system will have to be compatible with the manufacturing environment, sample the correct locations at the right size. and fulfill the FDA's requirement for batch information.

The FDA currently requires the manufacturer to show that the process area was in control only during the actual filling operation. Continuous monitoring of the whole filling process is preferred, but not required. Monitoring Is not required when the room is not in use. It is good practice to monitor during set-up to be assured of the quality of the air, but it is not required.

If the product or personnel are changed during a shift, a new monitoring program must be started. The particle counts taken during the morning do not reflect the operating conditions later in the day if a new product or new personnel have been introduced. Changing shifts would also require a new verification of the mom, even if the same product was being made. The monitoring procedure must occur during the actual filling operation, with the equipment and personnel operating in their typical modes. This approach makes good sense in a majority of cleanroom operations.

The monitoring of microorganisms in the product and in the environment is still the main focus for assuring product sterility: monitoring of particles - living or not-is useful as an additional indicator of control and as a "real-time" indicator of potential contamination.

Federal Standard 209D

Although the FDA uses Federal Standard 209D as a general basis for their guidelines. the FS-209D Is not an absolute set of rules for the pharmaceutical industry. One of the essential differences between the FDA's philosophy and FS-209D is that FS-209D allows averaging of counts across sample points. During batch filling, averaging data across multiple sampling points is not acceptable to the FDA. To keep proper records for a sterile fill area. the particle counts obtained at each critical point must be maintained as a separate record. The intent here is to monitor the air approaching the filling line, not to characterize the general particle level of the room.

The "Guideline on Sterile Drug Products produced by Aseptic Processing" states that the sample points should be within one foot of the fill line. The FDA will accept a greater distance if It can be shown that sampling within one foot will produce erroneous counts due to some processing condition. such as over spray from a highspeed filling needle or a powder filling operation.

Also In contrast to FS-209D. the "Guideline ... for Aseptic Monitoring" requires monitoring at 0. 5 microns: monitoring at other size ranges, for example 1.0 or 0.3 microns, must be well justified for the particular filling process. For aseptic areas. the counts at each sample point must be below 100 counts per cubic foot for all particles 0.5 microns and larger. Ibis generally equates to an FS 209D "Class 100." but the FDA specifically requires monitoring at "0.5 micron and larger" sizing. Using alternate size equivalents is not acceptable unless the manufacturer can justify an alternate size based on unique processing requirements.

Establishing Quality and Control of the Processing Area Using Trends

At a given sample point, repeated monitoring will provide a typical "baseline" of values during the manufacturing operation. "Alert" and "Action" levels can be set relative to the empirically determined baseline count. Using 1 -sigma and 3-sigma points derived from the baseline count might be appropriate. The FDA is comfortable with averaging data at a given sample site: that is. each and every sample does not have to be below the "Alert" level to continue the processing the batch. However, a consistent trend away from the baseline should be a signal that the area is out of control.

Selection of Sample Points

Within the sterile fill area. samples must be taken in the immediate vicinity of the fill heads, and preferably near where the containers enter the sterile area. In general. you should monitor at points where the product Is exposed to potential contamination and where an operator might frequently be present.

Outside of the sterile fill area or other "critical" areas. the FDA recommends frequent monitoring to ensure that the HVAC filtration system is in order, but the frequency of monitoring can be relaxed to between daily and once-per-week. Again. the intent of gathering this data is to show that the area is generally in control.

Tubing Extensions

Short extensions - up to a maximum of roughly 10 feet - from the sample point to the sensor are generally acceptable, assuming that the tubing has a minimum of turns or curves and that the curves have a generous radius. Due to the statistically low number of particles within a sample under "Class 100" conditions, it is best to limit the use of tubing, which causes some entrapment or fragmentation of particles. If the tubing must be longer than 10 feet. then the loss factor for that given tubing must be determined and a correction factor must be used to adjust the counts obtained during filling procedures. In general, the use of manifolds for sampling in extremely clean areas (e.g. Class 10) Is strongly discouraged by the FDA.

Saving Data

In general. for most sterile products. the particle count data should be saved for a period extending one year past the expiration date of the product. There is no restriction on the form of the data storage: it must simply be retrievable for inspection within a reasonable time.

Equipment Considerations

Sterile fill areas are often aseptically washed. However, particle counters must be protected from moisture. Since tubing length restrictions require the particle counter/sensor to be in the washed area, the equipment must be housed in a moisture and splashproof container. such as a NEMA 4X rated enclosure.

| Application Notes Menu | Met One Product Menu |

 
 

 

iProcesSmart Home Page of Industrial Process Equipment 
Model 228
HHPC-2
HHPC-6

Model 237

Model 2400
Model 3313

Accessories


 

SEARCH  INVENTORY



 

 

 

 


 
Click here to return to the main page! Home Model 228 ] [ HHPC-2 ] [ HHPC-6 ] [ Model 237 ] [ Model 2400 ] [ Model 3313 ] [ Accessories ]
 
   
Important Notice: By purchasing products from this Site, you acknowledge and agree that you have read and accept the Terms and Conditions of this online store. For more information email us at: info@iprocessmart.com